24 Performance Verification
YM6000 Service Manual
Pressure
Sensor
Accuracy
Test
Figure
8.
Pressure
Sensor
Accuracy
Test
1.
Connect
the
NIBP
cuff
hose
to
the
NIBP
connector
on
the
monitor’s
front
panel.
2.
Connect
the
other
end
of
the
NIBP
cuff
hose
to
the
NIBP
simulator.
3.
Rotate
the
trim
knob
to
select
NIBP
Test
in
the
service
menu,
and
then
press
the
trim
knob.
4.
Rotate
the
trim
knob
to
select
Pressure
Sensor
Accuracy
Test
,
and
then
press
the
trim
knob.
5.
Press
Select
button
on
the
simulator
until
simulator
displays
“Pressure
Source
Set
Test
Pressure”.
Adjust
pressure
on
the
simulator
for
250,
150,
50
and
0
mmHg.
6.
Press
Start
Pump
button
on
simulator.
The
simulator
will
begin
to
pressurize.
Allow
15~20
seconds
for
pressure
to
stabilize.
7.
Compare
the
pressure
of
the
simulator
with
the
pressure
displayed
by
the
monitor.
If
air
leakage
disables
accurate
comparison
of
the
above
pressures,
eliminate
the
cause
of
air
leakage.
8.
After
finishing
the
test,
press
the
trim
knob
to
select
Cancel
.
The
menu
box
will
disappear.
If
Cancel
is
selected
during
the
test
progressing,
the
test
will
stop
and
the
menu
box
will
disappear.
Pass/Fail
Results
The
difference
between
the
simulator’s
and
the
monitor’s
readings
are
as
follows:
Sphygmomanometer
Monitor’s
Readings
0
mmHg
±0
mmHg
50
mmHg
50±6
mmHg
100
mmHg
100±6
mmHg
200
mmHg
200±6
mmHg
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...