Troubleshooting 55
YM6000 Service Manual
Alarm
Messages
Check
Items
pressure
(C14)
a
possibility
that
standard
cuff
pressure
might
be
detected
wrongly
due
to
noises,
motion
artifact
or
external
vibration.
Check
whether
cuffs
are
not
wrapped
around
thick
clothing,
whether
the
patient
is
moving
and
whether
cuffs
are
free
from
outside
vibrations,
then,
measure
again.
When
it
occurs
in
the
initial
measurement
in
continuous
mode,
the
second
measurement
will
start
unless
Stop
button
is
pressed.
NIBP:
Cuff
irregular
pulses
(C15)
Blood
pressure
could
not
be
measured
because
oscillation
graph
was
not
normal.
There
is
a
possibility
that
motion
artifact
or
vibration
from
outside
might
interrupt
the
measurement.
Check
whether
the
patient
stays
still
and
cuffs
are
free
from
external
vibration,
then,
measure
again.
When
it
occurs
in
the
initial
measurement
in
continuous
mode,
the
second
measurement
is
continued
unless
Stop
button
is
pressed.
NIBP:
Cuff
motion
artifact
(C16)
Blood
pressure
could
not
be
measured
because
noises
interrupted
pulse
waveform
signal.
Check
for
motion
artifacts,
or
external
vibration
and
then,
measure
again.
When
it
occurs
in
the
initial
measurement
in
continuous
mode,
the
second
measurement
will
start
unless
Stop
button
is
pressed.
NIBP:
Cuff
time
‐
out
(C17)
Measurement
was
preventively
stopped
because
measurement
time
exceeded
160
seconds
(adult),
There
is
a
possibility
that
blood
pressure
might
be
repeatedly
measured
due
to
insufficient
pressurizing
caused
by
calcified
pseudohypertension.
NIBP:
Cuff
time
‐
out,
over
160
pulses
(C18)
Pulse
waveform
signal
more
than
160
beats
are
detected
during
measurement.
There
is
possibility
that
noises
might
interrupt
signal.
Motion
artifact
or
external
vibrations
possibly
affected
cuffs.
Check
for
patient
movement
and
if
the
cuff
is
free
from
stays
still
and
cuffs
are
free
from
outside
vibration,
then,
measure
again.
NIBP:
Cuff
pressure
failure
(C19)
Cuff
pressure
exceeded
more
than
300
mmHg
(adult)
during
measurement.
There
is
a
possibility
that
the
patient
moved
during
measurement
or
strong
pressure
from
outside
might
be
added
to
cuffs.
Considering
above,
measure
again.
NIBP:
Cuff
weak
pulse
(C20)
Amplitude
of
pulse
obtained
from
cuffs
are
too
weak.
This
error
possibly
occurs
when
cuffs
are
wrapped
around
loosely
in
ASO
patients
or
when
cuffs
are
wrapped
around
thick
clothing.
Wrap
cuffs
around
properly,
then,
measure
again.
NIBP:
Check
cuff,
hose
and
mode
(C21)
Patient
to
be
measured,
and
cuff
size
used,
do
not
match.
This
error
may
occur
if
the
blood
pressure
measurement
mode
setting
is
incorrect,
if
the
cuff
has
been
wrapped
tightly
in
the
adult
mode,
loosely
in
the
neonatal
mode
or
if
the
arm
has
been
bent
during
measurement.
Check
the
measurement
mode
setting
and
application
of
the
cuff,
and
measure
again.
NIBP:
Internal
error
(E03)
NIBP
module
error
BPM
pressure
sensor
fault.
Pump
operated
for
ten
seconds,
however
pressure
does
not
change.
Check
the
connection
of
the
cuff
hose.
NIBP:
Internal
error
(E07)
Reboot
the
monitor.
If
the
problem
persists,
cease
use
immediately
and
contact
qualified
service
personnel
or
your
local
supplier.
Содержание YM6000
Страница 6: ...iv Contents YM6000 Service Manual This page is intentionally left blank...
Страница 14: ...8 Introduction YM6000 Service Manual This page is intentionally left blank...
Страница 50: ...44 Service Menu and Factory Default Settings YM6000 Service Manual This page is intentionally left blank...
Страница 66: ...60 Troubleshooting YM6000 Service Manual This page is intentionally left blank...
Страница 82: ...76 Disassembly Guide YM6000 Service Manual This page is intentionally left blank...