Page 57 P/N 03521015 Rev U1
Bio-Medical Devices Intl
17171 Daimler Street Irvine, CA 92614 800-443-3842 www.maxair-systems.com
13. Charger
WARNING
The Single Chargers, 2600-01 (01432089) and 2600-02 (01432202), should only be used in an isolated area away from
patients and other activities, and away from flammable materials. Inspect the charger for damage before every use.
Do not use if damage is apparent or suspect.
A battery should be connected to a charger only until the Charger LED turns Green indicating a fully charged Battery.
When the Charger LED turns Green, the Battery should be disconnected from the Charger.
Refer to Section 28 for details regarding intermittent use and storage of batteries.
Intended Use
1. This Charger is designed for indoor use only and should not come into contact with water or excessive dust. To prevent
overheating the product should not be covered during use.
2. The mains socket should be easily accessible. In the event of operational error, the plug should be immediately removed
from the socket.
3. This Charger is designed for use with MAXAIR Lithium-Ion Batteries. For safety reasons, this Charger must be used only
for MAXAIR Batteries which have the right number of cells in series: Output voltage divided by 4.1V or 4.2V.
4. The Charger contains dangerous voltages and the cover should not be removed.
5.
All recommended maintenance work should be carried out by qualified personnel who can get assistance by contacting the
manufacturer’s agent.
6. A fuse protects the Charger against short circuiting and overloading.
7. This symbol
means that the charger is double insulated (Insulation Class II)
8. If the Charger is mounted in a vehicle it can only be used when the vehicle is not in use.
9. If the Charger is labeled “EN60601-1” and therefore it complies with the requirements of electro-medical equipment, it can
be used in hospital environments, etc.
10. The Charger has a plastic casing; avoid its coming into contact with oils, grease etc., as most types of plastic can be broken
down by chemicals and solvents.
Charging Instructions
1. Connect the Charger (single and/or
6-Gang) to an appropriate grounded
wall mains power source (120-240
VAC, 50-60 Hz) before connecting
to the Battery(ies). The Charger
green status LED should turn on.
2. Connect the Battery(ies) to the
Charger(s) by pushing the Charger
Cord Connector into the Battery
Connector Receptacle until fully
seated. The Charger LED should
change from green to red to indicate
charging. If the LED is Green after
being connected to the Battery for 10
seconds, the Battery is ready for use.
3. When charging is complete, the
Charger LED should change to
green. Disconnect Battery(ies) from
Charger(s) by pulling the Charger Cord
Connector from the Battery Connector
receptacle. The Battery(ies) is(are)
ready for use.
WARNING
The Charger has internal fuses which blow if a fault occurs in the charger. Additionally, the Charger is equipped with
a fuse switch which cuts off the unit in the case of a reverse polarity connection to the Battery. If a Charger fails,
contact Customer Service at 1-800-443-3842 for a Return Material Authorization (RMA).