CAUTION
Prior to each use, if any of the following issues are discovered for any system component(s), replace the
particular item(s) by following the assembly/disassembly procedures for the particular item(s).
•
Tears or Breaks.
•
Contamination from blood or other bodily fluids not safely removed by following approved
disinfection procedures.
•
Compromise between the DLC and FCC seal.
•
Damage or distortion to the filter cartridge gasket.
•
Filter is soiled or challenged with particulates such as to compromise its performance or cause the
yellow LED to be lighted.
•
Compromise between the filter cartridge and helmet seal.
•
Any other damage and threat to proper function.
5. General System Maintenance and Storage
NOTE
The complete MAXAIR CAPR System and all components and accessories should be stored indoors in a safe, clean
and secure environment at all times, protected from adverse environmental conditions, i.e. conditions that would be
considered incompatible with normal human working conditions without special equipment.
General System and Component Storage Environment
Temperature/ Humidity
Temperature: -20
◦
C to +40
◦
C
Maximum Humidity: 80% Relative Humidity.
Helmet Liner
•
If the Helmet Liner is loosened after repeated assembly/disassembly so as to compromise its attachment mechanism or
causes the Helmet mounting to be unstable and if there are any tears or breaks in the Liner, the Helmet Liner should be
replaced by following its assembly and disassembly procedure.
•
The Liner Power Cord Slot allows removal and attachment of the Liner without removal of the Power Cord.
Helmet
● If the Helmet is damaged or operating improperly, do not attempt repair. Contact Customer Service, 1-800-443-3842 for the
return procedure for evaluation and possible repair or replacement.
Helmet Power Cord
•
If the Helmet Power Cord connectors and cord insulation appear damaged in any manner, and if any cord wire is exposed,
replace with a cord in good working condition. Grasp the cord connector firmly and turn it until its notch is aligned with the
Helmet cord connector slot, then pull the cord connector straight up and away from the helmet. Replace with a new cord by
aligning the Cord connector notch with the Helmet connector slot, insert the Cord connector into the Helmet connector, and
turn the Cord towards the rear of the Helmet to secure it in place.
Filter Cartridge
● If the Filter is soiled or loaded (clogged) with particulate such as to compromise its performance or cause the Yellow LED to
be lighted, or if there are tears or breaks, or if there are compromises between the seal and the Helmet, or any other damage,
the Filter Cartridge should be replaced by following the assembly and disassembly procedures for the FCC and the Filter
Cartridge.