41
MagCore®
Plus II
Signal pathway:
Sensor → Sensor Cable → PCB → PLC → PCB → HMI
1.
Check if the SENSOR working properly:
Check PLC – X12 should light up when the sensor is ON and turn off when the
sensor is OFF:
Machine is turned on.
Sensor is working properly,
LED is on.
Place an object next to the
sensor → LED will turn off.
Remove an object → LED
will turn back on.
Checking Signal Propagation
I/O
IN
I/O
OUT
X0
Y Motor Encoder Signal(A) Y0
Y Motor Drive(A)
X1
Y Motor Encoder Signal(B) Yl
Y Motor Drive(B)
X2
Y Home Sensor Signal
Y2
V Motor Drive(A)
X3
Y3
V Motor Drive(B)
X4
V Motor Encoder Signal A
Y4
X Motor Drive(A)
X5
V Motor Encoder Signal B
Y5
X Motor Drive(B)
X6
V Home Sensor Signal
Y6
X7
UP Sensor Signal
Y7
X8
Y8
X9
Y9
X10 X Motor Encoder Signal(A) Y10
M DC Motor JOG
X11 X Motor Encoder Signal(B) Y11
UV Lamp Switch-Relay
X12 X Home Sensor Signal
Y12
RS232 Switch
X13 Down Sensor Signal
X14
X15
X16
X17
Door Sensor