37
MagCore®
Plus II
2. Detach the housing.
3. Loosen 3 screws on both sides of the machine.
After the screws are loosened Y-axis frame can be moved backward and forward.
4.
Put the tips into the T-rack and adjust the position of the frame until pistons align
with tips.
Tighten the screws back.
5.
Calibrate X and Y following the calibration procedure (p.13-15).