━ Heating Test
Select Heater Test button.
Use a digital thermometer to get more precise measurements. Put it into the separate
Heat Block Wells and press Next.
Select Test Temperature, press Start and wait.
Be aware that Screen response time for 90 degree test is slower!
Now the heater will heat up to 65° C or 90° C; check whether the temperature shown
on the thermometer is consistent with what is shown on the screen.
The acceptable temperature range is 65± 10° C or 90± 10° C.
To get accurate measurements go to RD mode and run the heater test. Highlighted
numbers are the precise temperature and should be very close to the temperature
shown by the thermometer.