47
MagCore®
Plus II
If cartridges misalign with the heating module, heater will push them up during the
X-axis movement.
To check, put the cartridges in the cartridge rack, but don't install the T-rack. Put
cartridge rack inside the machine, make sure that all the cartridges are properly placed
and run M-motor test – cartridges should not move.
1.
If M-axis has a problem and is not properly aligned with X-axis heating module will
push the cartridges up.
To check M-axis make sure the gap between heater and magnetic bar is the same
along the whole length of the module and is equal to the size of the cartridge, like
on the picture:
━ Cartridges alignment
━ M-axis Tolerance Misalignment