![MagCore Plus II Скачать руководство пользователя страница 60](http://html2.mh-extra.com/html/magcore/plus-ii/plus-ii_troubleshooting-manual_3403766060.webp)
58
www.rbcbioscience.com
R<4.11Ω
The heater is broken
R=0Ω
The heater is broken
R>4.11Ω
The module is aging.
The smaller R is the bigger power will be, it is better to replace the part.
2. Check if all SSR connections are tight:
3.
If connections are loose, tighten them and make sure no cables drop off.
4.
Measure the voltage at the SSR input.
SSR operation:
SSR
SSR LED
Heater
V in
On/Off
Blinking
On
Can't be
measured
Run heater test
Off
Off
Off
0V
When heater test is not running
On
On
On
12V
Heater test is running, D4.T value is large and stable