Refer to the picture below, switch cable
<X.ECD>+<X.HOME>+X motor line
with
cable
<V.ECD>+<V.HOME>+V motor line
Turn on the power, check if X axis is homing normally during initialization. If it is, X
motor and motor line have no problem, the most likely cause is the malfunction of
X axis driver on the motherboard. Replace the motherboard and test.