━ Activating All the Menu Buttons
Calibration
If some buttons of the Maintenance menu remain inactive after installation, switch off
the machine, close the door, turn the machine on and let it run automatic initialization.
After it stops making noise check if those buttons are activated.
This issue happens when PLC is new and it does not have the initial position values
written in. When machine initializes the values are being written into PLC. You will have
to do this once and they will stay in the memory.
Tap "Maintenance" → next page → "Advanced" (Password-1688).
Go to page "Teach position" and follow the steps below to calibrate.