Check if X12 works properly in I/O:
● When the sensor is on X12 will light up in yellow
● When the sensor is off X12 will stay white
2. Check cables
Check if the connection to the sensor is good, and if the connectors are not
damaged:
Check if the connections to PCB are good: cable CBL-0032 (green line) to X10
terminal. Check other connection:
Make sure all the
pins are connected
correctly.
Make sure this connection is
tight. Disconnect and connect the
cable back.
Make sure this connection
is tight. Disconnect and
connect the cable back.
1.
Unplug the cable, make sure all the
cables inside are well connected.
Connect the cable back following the
proper order (yellow, white, blue).
2.
If the LED is still off → change the cable
3.
If LED does not react to the objects
next to the sensor → change the cable
+ INFO
-
Wrong!
Correct
Maintenance-IO Check
Back
X00 Y Motor A CMD Pulse
X01 Y Motor B CMD Pulse
X02 Y Home Sensor
X03 Space (Reserve)
X04 V Motor A CMD Pulse
X05 V Motor B CMD Pulse
X06 V Home Sensor
X07 Up Sensor
X10 X Motor A CMD Pulse
X11 X Motor B CMD Pulse
X12 X Home Sensor
X13 Down Sensor
X14 Space (Reserve)
X15 Space (Reserve)
X16 Space (Reserve)
X17 Door Sensor
Y00 Y Motor Pulse
Y01 Y Motor Reverse
Y02 V Motor Pulse
Y03 V Motor Reverse
Y04 X Motor Pulse
Y05 X Motor Reverse
Y06 Space (Reserve)
Y07 Space (Reserve)
Y10 M A DC Motor JOG
Y11 UV Lamp Relay
Y12 Space (Reserve)
Y13 Space (Reserve)
M Axis Up
M Axis Dn
LED Lamp Off
R
G
B
W
Y10
Y13
Y11
Y12
IO Test Mode Off
X Motor
Y Motor
V Motor
25128
0
-1
25130
0
0
Encoder
CMD
07/13/2020 09:25:32
Содержание Plus II
Страница 1: ...MagCore PlusII Troubleshootingguide Ver 2020 1230 www rbcbioscience com...
Страница 8: ...6 www rbcbioscience com PCBConfiguration...
Страница 36: ...34 www rbcbioscience com AxisMisplacement AxisDirections...
Страница 61: ...59 MagCore PlusII Measureheatervoltage...
Страница 72: ...70 www rbcbioscience com 8 GotopageSystemStatusandchecktheProgramVer...