60
www.rbcbioscience.com
Software Update
When a new version is released we recommend updating all PLC, HMI and Program
at once to avoid error due to incompatibility of the software versions.
There is no particular sequence for software update, you can update HMI, PLC and
Program in any order.
For SW update you will need a Software Update Tool Set and make sure that your
computer has RS232-usb converter cable driver installed.
After Software Update (HMI, PLC and Program) rewrite X, Y and V values and speed
into the PLC. For X and Y copy existing value and save it. For V and Speed load default
values and save them.
one USB to RS232 adapter
for DVP ES/EX/EH
one RS232-PLC
programing cable
one USB to RS232
converter
OR
Prepare your Software Update Tools
Download PLC software update tool "EasyClient" from the link below and install.
https://download.rbcbioscience.com/MagCore%C2%AE/MagCore%C2%AE%20
Automated%20Nucleic%20Acid%20Extractor/MagCore%20SUPER/Software/
EasyDownload_v1.83/EasyClient.exe