48
www.rbcbioscience.com
2.
Make sure the heating wells of the cartridges are properly aligned with the holes of
the heating module, and nothing is loose anywhere along the M-axis.
3.
If the distance is not the same on two sides, loosen 4 screws:
4. Adjust the distance between heating rod and magnetic bar.
5. Tighten the screws back.