77
MagCore®
Plus II
5
Instrument is dropping tips
•
Y-axis is not calibrated properly;
•
There is too much grease on the piston tips;
•
X-rings and O-rings wore off and pistons need maintenance or replacement;
Loose axis coupling
1 Recalibrate Y-axis accordingly
2 Remove extra-grease and examine the condition of pistons
3 Check Y and V-axis coupling, tighten the nuts if needed
6
Heater is not warming up
•
Parts are wearing off;
•
PCB is broken and cannot control heating module
1 Check all the wires and connections to PLC (CBL_C006), PCB (RS232) and power supply, make sure
they are tight
2 Check heater resistance
3 Check resistance of the heater switch
4 Check if there is voltage at the input and output terminals of SSR
5 Check Pick IC : go to the System Status --> Pick IC If the number is 0279 Pick IC is functioning
properly, if it is 0000 Pick IC has to be replaced
7
X-Axis is not moving, triggering error 20
•
Internal spillage
1 Use large amount of water and grease on the rails and in the sliders to clean them; make sure the
grease gets inside the bearings
2 Clean the rails and run X- motor test
3 If there is still rust on the rails repeat until they are clean and lubricate them
8
Door mechanism touches cartridge rack
•
Improper transportation
Check the gap at the back of the X-axis to see if cartridge rack is properly aligned with the axis
9
The gap between HMI and the screen is too big
•
Improper transportation;
•
Loose screws
1 Move HMI frame forward
Содержание Plus II
Страница 1: ...MagCore PlusII Troubleshootingguide Ver 2020 1230 www rbcbioscience com...
Страница 8: ...6 www rbcbioscience com PCBConfiguration...
Страница 36: ...34 www rbcbioscience com AxisMisplacement AxisDirections...
Страница 61: ...59 MagCore PlusII Measureheatervoltage...
Страница 72: ...70 www rbcbioscience com 8 GotopageSystemStatusandchecktheProgramVer...