1
MagCore®
Plus II
Warning:
Always check packaging when receiving a new unit!
Let us know immediately if indicators are not in the initial condition (showing
that machine was tilted during transportation).
Perform the following and send us the result:
1. Take a picture of indicators.
2. Keep the machine switched off. Push X-axis inside and close the door.
3. Turn the machine on. It will run automatic initialization, let it run and when
machine stops making noise and nothing moves inside, go to Maintenance
and run piston test.
Notice:
If the problem you are experiencing is not covered by this guide, let us know:
fill in the Instrument malfunction form and send it to us along with the
instrument's serial number!
To see RBC training videos click or copy in your browser the link below:
MagCore Plus II Training Files
(https://drive.google.com/drive/folders/1ORi-jZ3MWz0-2S4b4-8wEud7sLbADD0-)
Содержание Plus II
Страница 1: ...MagCore PlusII Troubleshootingguide Ver 2020 1230 www rbcbioscience com...
Страница 8: ...6 www rbcbioscience com PCBConfiguration...
Страница 36: ...34 www rbcbioscience com AxisMisplacement AxisDirections...
Страница 61: ...59 MagCore PlusII Measureheatervoltage...
Страница 72: ...70 www rbcbioscience com 8 GotopageSystemStatusandchecktheProgramVer...