Chapter 1
_______________________________________________________________________________________
_________________________________________________________________________________________
BB6032-14
Photoacoustic Gas Monitor
LumaSense Technologies
1314i / 1412i / 1512 / 3434i
Page 14 of 81
IS 1314i/1412i/1512/3434i’s IP ADDRESS SET BY DHCP?
This determines whether the IP address of the Monitor is fixed or if the IP
address is set by DHCP.
DHCP is a server on the Ethernet network who automatically assigns IP
addresses to Ethernet devices on the network.
1314i/1412i/1512/3434i’s IP ADDRESS
If the IP address is not set by a DHCP server (as selected in the previous
setup) the IP address of the Monitor can be entered manually.
1314i/1412i/1512/3434i’s IP PORT NUMBER:
The IP port number where communication between the Monitor and host PC
goes through. The default port number is 23.
PROTECTED IP ADDRESS
This sets the IP address of the host PC, who is allowed to communicate with
the Gas Monitor.
If the value is 000.000.000.000 the any host PC is allowed to communicate
with the Gas Monitor.
CURRENT IP ADDRESS
This is shown if the IP address is set by the DHCP server. It tells which IP
address have been assigned to the Gas Monitor by the DHCP server.
1.3.2 The Homepage of the Gas Monitor
The homepage of the Monitor can be displayed if the Monitor is connected
to a Ethernet (TCP/IP) network.
Any web browser can be used to display the homepage of the Gas Monitor.
The IP address obtained from the TCP/IP set-up of the Monitor can be
entered in the address field of the web browser in order to display the
homepage.
The last measured gas values are displayed together with information about
the Gas Monitor, like the device serial number, if water- and cross-
compensation is performed.
Error and warning messages will also be displayed on the homepage.
The grey "Update Measurements" link can be used to update the homepage
with the last measurement results.