BB6032-14
Photoacoustic Gas Monitor
LumaSense Technologies
1314i / 1412i / 1512 / 3434i
Page 37 of 81
O_M? AL,N or O_M? AL,N,0
Returns the total numbers of alarm events for all
channels.
O_M? AL,N,1
Returns the total numbers of alarm events for channel 1.
O_M? AL,N,F_A Returns the total numbers of alarm events for Filter A.
O_M? AL or O_M? AL,0
Returns the total alarm log for all channels.
O_M? AL,1
Returns the total alarm log for channel 1.
O_M? AL,1,10 Returns the 10 latest alarm events for channel 1.
O_M? AL,F_A Returns the total alarm log for Filter A.
O_M? AL,F_A,10
Returns the 10 latest alarm events for Filter A.
The format is: date,time,channel,Gas,concentration
Eg.
2016-02-26,08:16:23, 9,Gas A,20.305E+03
2016-03-01,14:07:50, 1,Gas A,31.395E+03
2.3.25 OUTPUT_SETUP_DATA?
O_SE_D? reads out setup data from the Monitor in ASCII form. The
parameter block, which defines what is readout can be one of the following:
ALl
COMmunication
CONfiguration
ENvironment
FIlters
FOrmat
General