Chapter 4
_________________________________________________________________________________________
_________________________________________________________________________________________
BB6032-14
Photoacoustic Gas Monitor
LumaSense Technologies
1314i / 1412i / 1512i / 3434i
Page 67 of 81
Error Message
Description
OUTPUT DATA CANCELLED
The current output-message is interrupted
by a new input message.
PARAMETER EXPECTED
Syntax error. The parameter for the
message is missing.
PARAMETER OUT OF RANGE
The numerical parameter is illegal.
RECEIVE ERROR
A fundamental error in receiving the
message.
RELAY CONTROL NOT ENABLED The user is trying to control the Monitor’s
alarm relays, but access to the relays has
not been opened.
STRING DATA EXPECTED
Syntax error. Illegal data type received.
STRING LENGTH OUT OF
RANGE
The string is longer than allowed.
TOO MUCH DATA
Syntax error. There are more parameters
in the message than necessary. Otherwise,
the message is correct and will usually be
executed.
UNKNOWN HEADER OR
CHARACTER DATA
Syntax error. Unknown message header
and character data.
WARNING: CTS HANDSHAKE
CONFLICT
The Monitor receives a handshake signal on
the RS–232 CTS line without being setup
for using CTS. There is a handshake
conflict between the Monitor and the device
connected to the RS–232.
WARNING: DSR HANDSHAKE
CONFLICT
The Monitor receives a handshake signal on
the RS–232 DSR line without being set up
for using DSR. There is a handshake
conflict between the Monitor and the device
connected to the RS–232.
WARNING: MEMORY FULL
WARNING: MEMORY LOCATION
ALREADY USED
WARNING: MEMORY LOCATION
EMPTY
Errors in connection with the use of the
Background Memory.
WARNING: NO DATA FOR
There is insufficient data in the Monitor to
make a print.
WARNING: RS232 FRAMING
ERROR DATA LOST
The Monitor was receiving corrupted data
on the RS–232 RX line.
WARNING: RS232 OVERFLOW.
DATA LOST
The Monitor receives data on the RS–232
RX line faster than it can handle. The
internal buffer is full.
WARNING: RS232 OVERRUN.
DATA LOST
The Monitor receives data on the RS–232
RX line faster than it can handle.
WARNING: RS232 PARTIY
ERROR. DATA LOST
The Monitor receives data with illegal
parity.