Error Message
Description
WARNING: SETUP AND / OR
CALIBRATION CONFLICT
The Monitor is about to start a
measurement, but the setup of the Monitor
and the possibilities of the current
calibration are in conflict.
WARNING: TIMEOUT. DEVICE
NOT
READY
The Monitor wishes to send print/error-
/data-log. The transmission starts, but the
device stops reception of data for too long.
The device may be off-line.
The device may be too slow.
The device may no longer be properly
connected.
WARNING: X-ON/X-OFF
HANDSHAKE
CONFLICT
The Monitor receives the X-On/X-Off signal
via RS–232 without being set up to use X-
On/X-Off. There is a hand-shake conflict
between the Monitor and the device
connected to the RS–232.
WRONG DATA TYPE
Table 4.1 List of error messages at RS-232 interface