2.3.45 STOP_MEASUREMENT
STOP_M stops the current measurement or calibration task.
2.3.46 STORE_DEFAULT
STOR_D stores the current setup in non volatile memory. A full reset will
recall the saved setup as the current setup. Exists only as input command.
2.3.47 STORE_MEASUREMENT
STOR_M stores the contents of the display memory in the background
memory at a specified (NR_1) (1..10) location.
2.3.48 SYNCHRONIZE
SY? Reads out the pump synchronization condition. This message is for use
when an external pump (ex. A pump in the 1409) is connected to the
Monitor.
There are two possible outputs:
Yes
indicates that the Monitor pump will start only when a
synchronization command is received from the
controller. This is the required condition when an
external pump is connected to the Monitor. If the
synchronization condition is not set to Yes, and an
external pump is connected, then airway errors can
occur.
No
indicates that the Monitor pump is controlled by the
Monitor.
/***** PROTECTED 59 *****/ (see section
SY sets the pump synchronization condition:
Yes
the Monitor pump will start only when a
synchronization command is received from the
controller.
No
the pump will be controlled by the Monitor.
If a Yes or No parameter is not given, then the Monitor pump starts if the
Monitor is waiting for the Synchronize command.