Chapter 2
_________________________________________________________________________________________
_________________________________________________________________________________________
BB6032-14
Photoacoustic Gas Monitor
LumaSense Technologies
1314i / 1412i / 1512 / 3434i
Page 36 of 81
O_M? 1,”yyyy-mm-dd”,”hh:mm:ss”,”yyyy-mm-dd”,”hh:mm:ss”
Returns the measurements from channel 1 in the
display memory within the specified timeframe.
The monitor will return all measurements in the
display memory if no timeframe is specified.
Eg.
10, 9
2015-12-01,07:46:23,443.15E-03,-,-,-,-,5.9372E+03,0,0,0,0,0,0,0,1,0,0
The format is:
Number of samples total, number of samples in each transmission block
Date, time, Gas A, Gas B, Gas C, Gas D, Gas E, Gas W,P,O,W,B,F,A,R,V,J,S
Gas monitor marks:
P
Power up flag
O
Operational error flag
W
Warning flag
Gas marks:
B
Bad result, Monitor incapable of calculating the Gas Concentration
F
Filter Alignment error
A
Alarm limit exceeded
Multipoint sampler marks:
R
Reset done flag
V
Power failed flag
J
Job error flag
S
Software error flag
O_M? F_A,”yyyy-mm-dd”,”hh:mm:ss”,”yyyy-mm-dd”,”hh:mm:ss”
Returns the measurements from Filter A in the
display memory within the specified timeframe.
The monitor will return all measurements in the
display memory if no timeframe is specified.
Eg.
18,10
2015-12-01,07:37:06,421.16E-03,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,0,0,0,0,0,0,0,0,0,0
2015-12-01,07:37:33,-,417.19E-03,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,-,0,0,0,0,0,0,0,0,0,0
The format is:
Number of samples total, number of samples in each transmission block
Date,time,Ch.1,Ch.2,(etc. up to),Ch.24,P,O,W,B,F,A,R,V,J,S
The marks are the same as for the O_M? channel job.