Photoacoustic Gas Monitor
LumaSense Technologies
1314i / 1412i / 1512 / 3434i
Page 55 of 81
Single Point calibration of filter A with 345.34 ppm gas and with cross
calibration.
STArt_Calibration Calibration_Water_Span,18000
Calibration of water filter with 18000mg/m
3
water.
2.3.42 START_DELAYED_MEASUREMENT
S_D_M “d hh:mm” or S_D_M “dd hh:mm” starts a Monitoring task at
the specified time.
D ~ day, h ~ hour, m ~ minute.
2.3.43 START_MEASUREMENT
STA_M starts a monitoring task immediately.
2.3.44 STATUS_BYTE?
*STB? reads out the contents of the Status-Byte,
.The number
(NR_1) is in the range 0..255.
For example, if the Status Byte contained the binary number 00000001 and
you typed *STB? , the Monitor would return the decimal number 1, indicating
that the Data Ready bit was set. The contents of the Status Byte is shown
Bit no.
Bit name
1
2
3
4
5
6
7
8
Data ready
Reset activated
Command completed
Error
Busy
Abnormal
SRQ_line
Special event
Table 2.14 Contents of Status Byte