Chapter 2
_________________________________________________________________________________________
_________________________________________________________________________________________
BB6032-14
Photoacoustic Gas Monitor
LumaSense Technologies
1314i / 1412i / 1512 / 3434i
Page 38 of 81
MEAsurement
SYstem
Units
These names refer to the headings in the “setup tree”. All data below these
headings will be printed.
2.3.26 OUTPUT_SPECIAL_COMPILATION?
O_SP_C?
reads out specially compiled data from the Monitor
in ASCII form. This command requires two
parameters. The syntax is shown below.
Where type is one of the following:
APplication
Data for 7620 Application Software.
SAmple_DAta
Reads out data for all filters or a specified filter
O_SP_C?<type><filterid>,AL
Returns measurement data including chamber
temperature and air pressure.
The filter id is the same parameter used for
OUTPUT_CALIBRATION_DATA?
The SAmple_DAta output “readout” format for each sample consists of the
following:
[Filter Error, Alarm][10 characters per gas][3 error and warning bits][Event
no.][Time stamp]