Chapter 2
_________________________________________________________________________________________
_________________________________________________________________________________________
BB6032-14
Photoacoustic Gas Monitor
LumaSense Technologies
1314i / 1412i / 1512 / 3434i
Page 32 of 81
2.3.20 KEYBOARD_LOCKOUT
K_L?
indicates whether the front panel push-buttons of the
Monitor are locked.
K_L
followed by a parameter Y locks or N unlocks the front
panel push-buttons of the Monitor.
2.3.21 MEASURE_AIR_PRESSURE
The atmospheric air pressure is measured during reset, measurement,
calibration or by the command Measure_Air_Pressure.
M_A_P
starts an atmospheric air pressure measurement.
M_A_P?
reads out the atmospheric air pressure.
2.3.22 MEASUREMENT_STATE?
M_S?
indicates whether the Monitor is measuring. The Monitor’s
output can be Y, it is measuring, or N, it is not measuring.
2.3.23 OUTPUT_CALIBRATION_DATA?
O_C_D?
provides a readout of calibration data from the Monitor in
ASCII form. When issuing the interface message, the
command parameter can be one of the following:
ALl
Data for all filters
Filter_A
Data for filter A
Filter_B
Data for filter B
Filter_C
Data for filter C
Filter_D
Data for filter D
Filter_E
Data for filter E