Chapter 2
_________________________________________________________________________________________
_________________________________________________________________________________________
BB6032-14
Photoacoustic Gas Monitor
LumaSense Technologies
1314i / 1412i / 1512 / 3434i
Page 29 of 81
/***** PROTECTED 1 *****/
D_S selects the display buffer to be read out on the display.
For example, when:
D_S SY is typed, the Monitor will read out “SY” to the controller during
normal system, display.
2.3.13 EXECUTE_STATE?
EX_S? requests a status of the Monitor. The output from the Monitor can
come as one NR_1 or two NR_1s. The first NR_1 gives a general description
of the current status of the Monitor.
Output no.
(NR_1)
Name
Description
0
5
6
7
8
9
10
11
12
14
15
16
Display_Run
Meas_Clean_Up
Meas_Count_Down
Meas_Init_Mode
Meas_Run
Calib_Run
Memory_Run
Pause_Measurement
Print_Menu
Print_Run
Setup_Run
Test_Run
Normal idle state
Just after stop measurement
Waiting for time to run out
Measurement mode selected
Measurement running
Calibration running
Memory mode related
Monitor in Pause mode
Print selected but not started
Print in progress
Setup mode selected
Test mode
Table 2.8 The meaning of the output from EXecute_State?
The second NR_1 is used only if the first NR_1 indicates that the Monitor is
calibrating. It describes the calibration procedure. The key to the first
The second integer, given when the Monitor is calibrating, indicates one of
the following:
0
No error occurred
1
Warning(s) detected
2
Errors detected