2.3.14 EXTENDED_CONTROL
E_C is used to gain access to a protected message. The Extended_Control
message must be send prior to and every time a protected message is sent.
For example, the SYNCHRONIZE message is a “protected 59” message.
E_C 59 must be send Prior to sending the message. i.e.
E_C 59
SY
These messages send the synchronization to the Monitor.
Note: Protected Messages in this manual are identified with
/***** protected 59*****/(see section
2.3.15 IDENTIFICATION?
*IDN? gives the identification of the instrument in the following format:
B&K,1314,ssss,VPxxxx
B&K,1412,ssss,VPxxxx
or
B&K,1512,ssss,VPxxxx
where:
ssss string of max 10 characters containing the serial number.
xxxx software revision number.
If a 1409 Multipoint Sampler is connected to the Gas Monitor, the number
of channels will be added to the IDN respond, given the following format for
a 1409-12, as an example.
B&K,1412,ssss,VPxxxx,12