Perform a Clear All Procedure
In some cases it may be necessary to perform a Clear All procedure and reset all of the
program settings back to the factory default. This is usually done by a factory trained
technician that is upgrading software, transferring a board from unit to unit, or if a
memory
clear is needed. It is recommended that you save a configuration file of the
current program settings or copy all of the settings to a notepad before performing a
clear all if you plan to preprogram the unit with the same information.
Follow these steps to perform a clear all:
1. Enter the calibration mode by loosening the calibration bolt on the side of the LCR.iQ until
the
Main
Menu
is displayed on the screen.
2. Navigate to the
Setup Menu
and press
OK
.
3. Navigate to the
Security
option and press
OK
.
4. Navigate to
Reset Register Fields
option and press
OK
.
5. Select
Clear
All
from the list box that is displayed and press
OK
.
6. A prompt will appear: “Are you sure you want to clear all settings? All settings will revert
back to factory defaults including custom settings.”
7. To proceed with the clear all, press the
Yes
function key and the clear all process will
begin.
8. Once complete, all programmable fields (except Time and Date) will be reset to factory
default settings.
9. Reprogram the unit as appropriate.