during the multi-point calibration run. This action will store the point and flow rate, then
calculate the percent error to be applied to the selected point.
Point
– A list box for selecting the calibration point for the multi-point calibration. The LCR.iQ
is able to perform a multi-point calibration on as many as 16 separate points.
Options:
·
1-16
Flow Rate
– A numeric text field for entering the actual flow rate at which a multi-point
calibration run was made.
% Error
– A numeric field that is used when the LCR.iQ calculates the percent error for a
multi-point calibration point. The percent error field can also be manually entered for meter
systems that have a factory supplied accuracy curve.
Security
Security settings are for setting up and configuring the user level security, and also access to
the
Clear All
process, software upgrades, Audit Trails and Reset Register Fields feature.
Setting the security level to
Locked
will prevent an operator from accessing fields that the
owner may want to restrict.