Section 6
Interfacing
6-15
6.6 Using the 6262-03 Pressure Transducer
The 6262-03 is located inside the LI-6262 case, and senses pressure from a
manifold on the optical cell. Because the calibration function used in
LI-COR gas analyzers will correct for pressure changes if the pressure in the
optical cell is known, the 6262-03 allows for automatic pressure corrections
to be performed.
During installation, an adapter board is connected to the analyzer’s digital
board. The adapter board sends the transducer mV output to the Auxiliary
terminal (#15) on the terminal strip (Figure 6-6). Because the mV
output
of
the transducer is connected to the Auxiliary terminal on the terminal strip,
the signal can be measured by connecting a voltmeter or datalogger to the
Aux
and
Sig Gnd
terminals. Pressure can then be calculated externally, if
desired. Internal computations are performed provided FCTs 71, 72, 73, and
75 are set correctly (see
Software Considerations
below).
Digital
Board
Adapter
Board
Power
Supply
Pressure
Transducer
Terminal Strip
AUX INPUT terminal #15 = Pressure transducer mV output
AUX INPU
T
1 2
3
4
5
6
7
8
9 10 11 12 13 14 15 16
Figure 6-6. Block diagram of electrical connections.
Software Considerations
The pressure value (kPa) used in the calculation of CO
2
is stored at channel
code 43. Pressure can be entered as a constant at FCT 77, or it can be
calculated from the 6262-03 mV signal provided to the Aux Input. If you
are using the 6262-03, the Auxiliary Destination (FCT 73)
must
be specified