Section 4
Calibration
4-7
2.
Remove the 8 screws from the top and bottom of the 6262-04 cover.
3.
Attach the 6262-04 cover to the analyzer using the 2 screws from the
mounting clips. The screws are inserted into the holes where the
mounting clips were previously attached.
4.
Attach the 6262-04 to its cover by replacing the 8 screws.
5.
Screw the soda lime/desiccant tube mounting clips into the rear face
plate of the 6262-04. Attach the soda lime/desiccant tube.
6.
Plug the battery cable from the 6262-04 into the battery jack on the
analyzer.
7.
Attach the hoses:
Mg(ClO
4
)
2
→
REFERENCE IN
fitting (LI-6262)
Soda Lime
→
TO SCRUB
fitting (6262-04)
TO CHOPPER
fitting (LI-6262)
→
REFERENCE OUT
fitting
(LI-6262). (This tube is threaded through the 6262-04).
FROM CHOPPER
fitting (6262-04)
→
FROM CHOPPER
fitting
(LI-6262)
The finished assembly will look similar to that shown in Figure 4-3
below.
8.
Plug battery into the battery jack on the 6262-04, or connect a power
cord to the LI-6262 if using AC power.
Important Note:
The 6262-04
Reference Pump must be turned off independently; it will not shut off
when the LI-6262 is powered off.
9.
Turn on the pump.