![LI-COR li-6262 Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/li-cor/li-6262/li-6262_instruction-manual_1906652071.webp)
FCT 70-71
5-26
System Software
70 Set Miscellaneous
70 ** MISC **
↑
,
↓
,
↵
Enters the menu at function 70.
The 7x functions include those necessary for using the auxiliary input
channel (#15 on the back panel terminal strip). The input signal is
transformed by
Y = A + Bx + Cx
2
where x is the signal in mV, and the coefficients A, B, and C are entered in
FCTs 71, 72, and 75. The value Y can be treated as pressure (FCT 43),
reference CO
2
(FCT 29) or reference H
2
O (FCT 39), according to FCT 73.
71 Auxiliary A (Offset)
Aux A=
0.0
NEW= __
Auxiliary A (Offset) value.
*71<A>