Section 7
7-6
Maintenance
7.5 Fan Filter
The external fan filter should be cleaned (rinsed out) or replaced as needed.
The filter sits in a recessed cavity in the instrument case and can easily be
pulled out with your fingers. Two additional external air filters are included
in your spare parts kit (LI-COR part # 6560-159).
7.6 Fuses
There are two fuse holders on the back panel. The 1/2 amp slow blow fuse
is for 110 VAC line voltage and the 1/4 amp slow blow fuse is for 220 VAC
line voltage. The 2 amp fast blow fuse is for the external batteries. If the
LI-6262 fails to turn on, check the fuse for the power source you are using
(battery or AC). If the analyzer continually blows fuses it is in need of
repair.
NOTE
: At 50 Hz, 220 VAC, it may be necessary to use the 1/2 amp
slow blow fuse due to near saturation of the power transformer core.