![LI-COR li-6262 Скачать руководство пользователя страница 51](http://html.mh-extra.com/html/li-cor/li-6262/li-6262_instruction-manual_1906652051.webp)
FCT 04
5-6
System Software
04 Set Displays
Function 04 performs functions 91-99.
The LCD display has two lines, and each line can be configured to show any
channel code 21-29 (CO
2
), 31-39 (H
2
O), or 41-47 (MISC). As many as nine
two-line display configurations can be defined by the Set Displays routine.
Each display is identified by a number 1 through 9, and can be viewed by
pressing the appropriate number
without
pressing
FCT
.
NOTE:
Displays 1-6 (FCT 91-96) default to the configurations shown
below at initial power-up or system reset.
Function Codes
Display
(Top/Bottom Lines)
Description
1
21
CO
2
mV (non-linear)
22
CO
2
µ
mol/mol absolute
2
23
CO
2
µ
mol/mol differential
29
CO
2
µ
mol/mol reference
3
41
Temperature mV
42
Temperature
°
C
4
31
H
2
O mV (non-linear)
32
H
2
O mmol/mol absolute
5
33
H
2
O mmol/mol differential
39
H
2
O mmol/mol reference
6
43
Barometric Pressure (kPa)
44
Auxiliary mV
Set the displays as follows:
Press
FCT
, followed by 04. The display will show:
1 Top=21
NEW= __
Shows default channel code for
top of display 1. Enter new code
21-29, 31-39, or 41-47.
1 Bottom=22
NEW= __
Default channel code for bottom
of display 1.
Prompts to configure displays 2 through 9 follow. Press
FCT
to quit.