![LI-COR li-6262 Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/li-cor/li-6262/li-6262_instruction-manual_1906652015.webp)
Section 2
2-6
Setup & Operation
Prepare a soda lime/desiccant tube, as described in Section 7.4, and attach to
the
SCRUBBER
holder on the back panel.
Differential Mode Operation
●
Orient the scrubber tube so that the soda lime is on the bottom and the
magnesium perchlorate is on top.
●
Attach the bottom hose (nearest the soda lime) to the
FROM CHOPPER
fitting.
●
Attach the top hose (nearest the perchlorate) to the
TO CHOPPER
fitting
(Figure 2-1).
Figure 2-1. Hose connections for operation in differential mode.
IMPORTANT
:
Attach the hoses as shown in Figure 2-2. The hose
nearest the soda lime is attached to the
FROM
CHOPPER
fitting, and
the hose near the magnesium perchlorate is attached to the
T O
CHOPPER
fitting. This ensures that CO
2
is purged first, and then water
vapor. If this sequence is reversed, the air will pick up water vapor
from the soda lime before being returned to the chopper housing.