FCT 18-19
5-18
System Software
18 Reset Timer
Press
FCT 18
to reset the Timer function. The Timer (FCT 45)will begin
counting again at zero.
*18
19 Full Status
Outputs the full status of the analyzer to the RS-232C port, in a format that
can be read directly back into the analyzer, if desired. If you are accessing
this function remotely, you can specify selected information to be sent by
entering the optional parameter. For example, entering
*191
would send only the *01 (CO
2
Cal) information, and
*1913
would send only the *13 RS-232C print list. *19 will produce the entire list.
The entire list will appear something like this:
*01 3.5970E+01, 1.7913E+04, 1.4330E-01, 9.5609E-06, 7.8293E-
09,-1.1040E-12, 7.5366E-17, 0.0000E+00,0
*02 4.1340E+01, 1.5409E+04, 6.3281E-03, 3.1059E-06, 1.1238E-
12, 0.0000E+00, 0.0000E+00, 1.0000E+00
*05 0, 0.0000E+00, 0.0000E+00
*06 0, 0.0000E+00, 0.0000E+00
*07 1
*08 1, 0.0000E+00, 1.0000E+00
*08 2, 0.0000E+00, 1.0000E+00
*09 0,1,1
*13
*15 0
*71 5.9071E+01
*72 1.5360E-02
*73 43