FCT 77-78
5-30
System Software
77 Pressure
P(kPa)=
0.0
NEW= __
Barometric Pressure in kPa.
Barometric pressure
must
be entered correctly for CO
2
or H
2
O concen-
trations to be calculated correctly.
NOTE:
Barometric pressure may also be input in Setup command 03.
Pressure may also be input continuously using the 6262-03 Pressure
Transducer from LI-COR, or through connector #15 on the back panel
terminal strip, as described in section 6.4.
*77X
(where x is the barometric pressure in kPa).
78 Vapor Correction a
VpCrr A= 1.5
NEW= __
Vapor correction value a.
Vapor correction coefficient a is factory-entered (a =
∼
1.5). See Section 3.5
for a description of the vapor correction. This value is only used when
Vapor Correct (FCT 76) is in State 1 or 2.
*78a
(where a is the value of vapor correction a).