Section 6
Interfacing
6-13
6.4 Auxiliary Channel Inputs
Terminal 15 is connected to the auxiliary channel which may be used to
input an analog signal from a user-supplied external sensor. CO
2
and H
2
O
references may be input through this channel, as well as barometric pressure.
Use FCT 73 to designate which channel (29, 39, or 43) the analog signal will
be sent to. Figure 6-4 shows how to connect a sensor to the terminal strip on
the back panel of the LI-6262.
N O T E
: If the 6262-03 Pressure Transducer is installed in your
analyzer, the auxiliary channel input is no longer available; the raw mV
reading from the transducer can be
measured
at this terminal, however.
Please see Section 6.6 for a complete discussion of the 6262-03.
●
Connect the positive input from the sensor to the terminal labeled
AUX
.
●
Connect the negative input to the terminal labeled
SIG GND
.
The auxiliary channel has an input voltage range of
±
4.096V.
DAC1 5V
DAC1 100mV
DAC1 20mA
DAC2 5V
DAC2 100mV
DAC2 20mA
SIG GND
SIG GND
CO
2
0.1S
CO
2
1S
H
2
O 0.1S
H
2
O 1S
SIG GND
TEMP 5V
AUX
CHASSIS GND
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
EXTERNAL
SENSOR
Figure 6-4. Connection of an external sensor for analog signal input.