![LI-COR li-6262 Скачать руководство пользователя страница 72](http://html.mh-extra.com/html/li-cor/li-6262/li-6262_instruction-manual_1906652072.webp)
FCT 72-73, 75
System Software
5-27
72 Auxiliary B
Aux B=
0.0
NEW= __
Auxiliary B value.
*72<B>
73 Auxiliary Destination
Aux Dest=0
NEW= __
Destination channel of auxiliary
input signal.
The auxiliary channel input may be sent to channel codes 29 (CO
2
reference), 39 (H
2
O reference), or 43 (pressure). Entering any other code at
the
NEW=
prompt sets the destination to 0 and deactivates the input channel.
Setting this destination to 29, 39, or 43 causes the values of FCT 59, 68, or
77, respectively, to be ignored.
*73X
(where x is the destination channel 29, 39, 43, or <null>).
75 Auxiliary C
Aux C=
0.0
NEW= __
Auxiliary C value.
*75<C>