![LI-COR li-6262 Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/li-cor/li-6262/li-6262_instruction-manual_1906652016.webp)
Section 2
Setup & Operation
2-7
To chopper
From chopper
Soda Lime
Fiberglass Wool
Magnesium
Perchlorate
Figure 2-2. Hose attachments to external soda lime/desiccant tube assembly.
Absolute Mode Operation
IMPORTANT
: Air flow created by movement of the chopping shutter
disc is sufficient to purge the reference cell of CO
2
, but is not sufficient
to remove all of the water vapor in the reference cell. For this reason an
auxiliary pump such as the 6262-04 must be used when operating in
absolute mode. See Section 4.4 for more information.
●
Orient the scrubber tube so that the soda lime is on the bottom and the
magnesium perchlorate is on the top.
●
Attach the bottom hose (nearest the soda lime) to the
FROM CHOPPER
fitting. An auxiliary pump should be placed between the
F R O M
CHOPPER
fitting and the bottom hose of the scrubber tube.
Make sure
the pump is oriented to pump in the direction indicated in Figure 2-3
.
●
Attach the top hose (nearest the perchlorate) to the
REFERENCE IN
fitting.
●
Attach a jumper hose between the
REFERENCE OUT
and
TO
CHOPPER
fittings.