FCT 01
5-4
System Software
01 CO
2
Calibration
Function 01 is actually a group of functions; executing FCT 01 performs
functions 51-59 and 76.
The LI-6262 is calibrated using a 5th order calibration polynomial of the
form:
µ
mol/mol = A(mV) + B(mV)
2
+ C(mV)
3
+ D(mV)
4
+ E(mV)
5
. The
calibration coefficients (T,K,A,B,C,D,E) are entered into the LI-6262 at the
factory, and are shown on the calibration sheet.
CO
2
reference mole fraction (
µ
mol/mol) and the choice of water correction
depend upon your system setup. See functions 59 and 76 for complete
descriptions.
Check the calibration sheet to verify that the correct values are entered.
Prompt
Description
Related Function
C2 T=
Calibration Temperature
51
C2 K=
Calibration Constant K
52
C2 A=
Calibration Coefficient A
53
C2 B=
Calibration Coefficient B
54
C2 C=
Calibration Coefficient C
55
C2 D=
Calibration Coefficient D
56
C2 E=
Calibration Coefficient E
57
C2 R=
CO
2
Reference
µ
mol/mol
59
BandBrd, Dil
→
REF
Water correction selection
76
*01T,K,A,B,C,D,E,REF,VAPOR
FLAG