![LI-COR li-6262 Скачать руководство пользователя страница 73](http://html.mh-extra.com/html/li-cor/li-6262/li-6262_instruction-manual_1906652073.webp)
FCT 74
5-28
System Software
74 Average Time
Avg(s)=0
NEW= __
Signal averaging time.
Sets the software signal averaging time, from 0 to 30 seconds. Increasing
the average time decreases the noise level in any displayed channel.
The digital signal displayed is averaged over the entire time span set, after
conversion from the analog signal. This results in a significant decrease in
the apparent noise level, but slows the response time, as well.
This average time affects all computed values of CO
2
(FCTs 22 to 27), and
H
2
O (FCTs 32-38), whether displayed, or output to a DAC channel, or
output by RS-232C. It does not affect the raw mV signals on the back
terminal strip.
Setting Average Time to zero will turn off all digital averaging, and the
displayed or output values of CO
2
and H
2
O will reflect the hardware
response time.
*74X
(where x is any value from 0-30 seconds).