FCT 50-59
5-24
System Software
50 CO
2
Calibration
50 **CO2 CAL**
↑
,
↓
,
↵
Enters menu at function 50.
51-59 CO
2
Calibration Stack
The CO
2
calibration stack (FCT 51-57) contains individual CO
2
calibration
coefficients, as well as the prompt to enter the CO
2
reference in
µ
mol/mol
(FCT 59).
C2 R=0
NEW= __
Enter CO2 reference in
mmol/mol.
Code
Prompt
Description
51
C2 T=
Calibration Temperature
52
C2 K=
Calibration Constant K
53
C2 A=
Calibration Coefficient A
54
C2 B=
Calibration Coefficient B
55
C2 C=
Calibration Coefficient C
56
C2 D=
Calibration Coefficient D
57
C2 E=
Calibration Coefficient E
59
C2 R=
CO
2
Reference
µ
mol/mol
The CO
2
reference value (FCT 59) should reflect the dilution by water
vapor, if the reference is not dry. For example, if the reference gas contains
300
µ
mol/mol CO
2
when dry, and 0.5 kPa of water vapor is added before it
enters the reference cell, then C2 R = 300(1 - 0.5/P), where P is the local
barometric pressure.