Section 4
Calibration
4-13
CO
2
/H
2
O
Scrub
CO
2
/H
2
O
Scrub
Swap
Diff.
Zero
Aux.
Ports
LI-670
Sample
In
Reference
In
Sample
Ref.
Scrub both cells.
Absolute mode. Set span.
5.
Turn OFF the
S A M P L E
C O
2
/H
2
O Scrub
switch. Leave the
REFERENCE CO
2
/H
2
O Scrub
switch ON. Adjust the flow as needed.
6.
Set the LI-6262 span when the displayed reading is steady.
CO
2
/H
2
O
Scrub
CO
2
/H
2
O
Scrub
Swap
Diff.
Zero
Aux.
Ports
LI-670
Sample
In
Reference
In
Sample
Ref.
Scrub only the reference cell.
The LI-6262 is now ready for use in absolute mode. Differential mode
operation requires three additional steps.
Differential mode. Set zero.
7.
Perform steps 1 through 6, and then turn OFF the
R E F E R E N C E
CO
2
/H
2
O Scrub
switch. Adjust the flow to near that which will be used
during measurements.
8.
Adjust the LI-6262 to read zero when the displayed mole fraction is
steady. Only a small adjustment should be necessary (about 5 ppm)
when switching from zero to 350 ppm in the reference cell.
9.
Enter the span gas CO
2
mole fraction in FCT 59, or the span gas H
2
O
mole fraction in FCT 68 in the LI-COR analyzer. Read differential H
2
O
or CO
2
on the analyzer display. Check the span by scrubbing the sample
cell and looking for zero on FCT 22 or FCT 32.