FlexLine, 32
Operation
Setup step-by-step
1. Extend the tripod legs to allow for a comfortable working posture. Position the
tripod over the marked ground point, centring it as best as possible.
2. Fasten the tribrach and instrument onto the tripod.
3. Turn on the instrument, and, if tilt correction is set to 1- or 2-axis, the laser
plummet will be activated automatically, and the
Level/Plummet
screen appears.
Otherwise, press
FNC
from within any application and select
Level/Plummet
.
4. Move the tripod legs (1) and use the tribrach footscrews (6) to center the
plummet (4) over the ground point.
5. Adjust the tripod legs (5) to level the circular level (7).
6. By using the electronic level, turn the tribrach footscrews (6) to precisely level
the instrument. Refer to "Level up with the electronic level step-by-step".
7. Center the instrument precisely over the ground point by shifting the tribrach on
the tripod plate (2).
8. Repeat steps 6. and 7. until the required accuracy is achieved.
2
6
7
5
4
5
5
1
3
1
1
TSOX_013