Check & Adjust
FlexLine, 243
Check and adjust
step-by-step
1. Level the instrument with the electronic level. Refer to "3 Operation"- "Level up
with the electronic level step-by-step".
3. Press
REC
to measure to the target point.
2.
Aim at a point approximately 100 m from the
instrument which is at least 27° (30 gon) above
or beneath the horizontal plane.
V=90°
+ 27°
- 27°
TSOX_066