Personnel:
n
Specialized personnel
1.
Select the
Modules
Through-flow control
Control parameter
menu
items.
2.
Scroll to a control parameter and select it with the Enter key.
An entry window opens. You can now change the numerical value
with the softkeys. The values displayed for
Min:
and
Max:
specify
the possible entry range.
3.
Press enter to confirm the new value.
Table 29: Standard settings
Control parameters
Meaning
Basic setup (Default)
Valve Kp
Proportional compo-
nent Valve controller
5.0
Valve Ki
Integrative component
Valve controller
1.25
Pump Kp
Proportional compo-
nent Pump controller
0.2
Pump Ki
Integrative component
Pump controller
0.1
6.4 Standard control parameters
The thermal mass and the thermal connection of the application to the
thermostat may make it necessary to adjust the control parameters.
Only modify the control parameters if you possess adequate knowl-
edge of control system engineering.
V4
Integral process thermostats with flow control unit
84 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...