4.7.5 Protocol of the interface
Note the following instructions:
n
The command from the computer must be made with a CR, CRLF, or
LFCR.
n
The response from the thermostatic circulator is always made with a
CRLF.
n
After each command sent to the thermostat, it is necessary to wait
for the reply before sending another command. This ensures that the
sequencing of inquiries and answers is clear.
CR = Carriage Return (Hex: 0D); LF = Line Feed (Hex: 0A)
Table 6: Example with set point transfer of 30.5 °C to the thermostatic
circulator
Computer
Temperature control device
"OUT_SP_00_30.5"CRLF
è
ç
"OK"CRLF
4.7.6 Read commands
The interface module recognizes the following read commands, which you
can use to retrieve operating data of the constant temperature equipment.
Table 7: Temperature
ID Function
Unit, resolu-
tion
Command
2 Temperature set point
[°C]
IN_SP_00
3 Outflow temperature
[°C], 0.01 °C
IN_PV_00
4 Outflow temperature
[°C], 0.001 °C IN_PV_10
5 Controlled temperature (internal / external Pt / external analog / external
serial)
[°C]
IN_PV_01
7 External temperature T
E
(Pt100)
[°C]
IN_PV_03
8 External temperature T
E
(analog input)
[°C]
IN_PV_04
14 External temperature T
E
(Pt100)
[°C], 0.001 °C IN_PV_13
25 Overtemperature turn off point T_Max
[°C]
IN_SP_03
27 Limitation of outflow temperature TiH (upper limit)
[°C]
IN_SP_04
29 Limitation of outflow temperature TiH (lower limit)
[°C]
IN_SP_05
V4
Integral process thermostats with flow control unit
45 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...