3.
The following options are available:
n
[Status]
Select the option [Start] to start the program.
Once the program has started, you can press [Hold] to pause
it.
A paused program can be continued by pressing [Continue].
Select the option [Stop] to end the program.
n
[Edit]
n
[Cycles]
Here you enter the number of replays of the selected pro-
gram.
4.
Select the
Edit
menu item.
The program appears on the display and you can now edit it.
You can stop the programmer by pressing the [STOP] softkey. After
pressing the [START]softkey, the programmer continues to run in the previ-
ously selected mode (pause or active mode).
100 temperature-time programs can be saved. A maximum of 250
freely programmable segments can be used per program.
The programmer encoder can be controlled or modified using the
timer.
Setting
Description
No.
Program segment number
Tend
End temperature that should be reached
hh
Time in hours (hh) by which the specified tem-
perature should be reached
:mm
Time in minutes (: mm) in which the specified tem-
perature should be reached
Tolerance
Tolerance defines how close the temperature
should be to the set temperature before the next
segment is processed.
0.0 means that there is no tolerance. This means
that the program starts the next temperature after
the preset time, even if the output temperature has
not yet been reached.
Fig. 64: Program 1
Available settings
Fig. 65: In the program editor
V4
Integral process thermostats with flow control unit
105 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...