Application
Unit
IN 550 XTW FC
IN 950 XTW FC
IN 1850 XTW FC
Recommended cooling water pres-
sure difference
bar
3
3
3
Minimum cooling water pressure
difference*
bar
0.8
0.8
0.8
Maximum cooling water pressure
difference*
bar
10
10
10
Cooling water flow rate
l/min
8.0
20.0
26.0
* only approved for a max. cooling water temperature of 20 °C
Table 51: Integral XTW FC with two-stage water-cooled cooling unit
Application
Unit
IN 590 XTW FC
IN 1590 XTW FC
Cooling output at
20°C and pump level 8
kW
4.50
18.50
10°C and pump level 8
kW
4.45
15.00
0°C and pump level 8
kW
4.40
11.50
-10°C and pump level 8
kW
4.40
8.70
-20°C and pump level 4
kW
4.60
8.50
-30°C and pump level 4
kW
4.60
8.50
Heat discharge of the refrigerating
machine
---
Water
Water
Minimum/maximum cooling water
temperature
°C
10/30
10/30
Recommended cooling water tem-
perature
°C
15
15
Recommended cooling water pres-
sure difference
bar
3
3
Minimum cooling water pressure
difference*
bar
0.8
0.8
Maximum cooling water pressure
difference*
bar
10
10
Cooling water flow rate
l/min
15.8
26.0
* only approved for a max. cooling water temperature of 20 °C
13.5
Refrigerant and filling weight
The device contains fluorinated greenhouse gases.
V4
Integral process thermostats with flow control unit
155 / 171
Содержание IN 550 XT
Страница 61: ...5 3 2 Menu structure Fig 26 Menu Part 1 V4 Integral process thermostats with flow control unit 61 171...
Страница 157: ...14 Declaration of Conformity and certificates V4 Integral process thermostats with flow control unit 157 171...
Страница 158: ...Fig 85 Declaration of Conformity V4 Integral process thermostats with flow control unit 158 171...
Страница 159: ...Fig 86 Declaration of Conformity V4 Integral process thermostats with flow control unit 159 171...
Страница 161: ...V4 Integral process thermostats with flow control unit 161 171...
Страница 162: ...V4 Integral process thermostats with flow control unit 162 171...
Страница 171: ...Y Year of construction 23 V4 Integral process thermostats with flow control unit 171 171...